Electrodiagnostic studies are an important tool in diseases that affect both the peripheral and central nervous system. Neuroconduction studies are non-invasive medical procedures that help evaluate neuromuscular diseases because they provide a physiological assessment of the peripheral nerve, muscle, neuromuscular junction, dorsal root ganglion neuron and anterior horn neuron. Motor nerve conduction studies assess the motor axons by selectively recording the muscle response to nerve stimulation. Sensory nerve conduction studies evaluate sensory axons by recording directly from the peripheral nerves (
Most of the normality parameters for nerve conduction and late response studies used for the population of Hospital Universitario San Ignacio and, in general, for Latin American countries were extrapolated from Buschbacher studies of 1999 (
The values of each record vary according to each laboratory, depending on factors such as height above sea level, room temperature or patient’s height. From this we see the importance of having normality parameters for nerve conduction and late response studies for each electrophysiology and electrodiagnostic laboratory.
The aim of this study is to find the normality values for motor nerve conductions of the median, ulnar, tibial and peroneal nerves, sensory nerve conductions of the median, ulnar and sural nerves, and normality values for the F-wave and H-reflex latencies for the Electrodiagnostic Laboratory of Hospital Universitario San Ignacio.
Descriptive cross-sectional study conducted at the Neurophysiology and Electrodiagnostic Laboratory of Hospital Universitario San Ignacio, Bogotá. By non-probabilistic sampling and according to the estimates of the parameters found, the minimum sample size required was 153 records, considering a final precision of 10 % in the estimation of the mean and its measures of dispersion and central tendency to generate reference values for motor and sensory nerve conduction. A total of 77 healthy adult patients were included, for whom 154 records were collected. The age of inclusion was 18 to 65 years of age.
Healthy companions of hospital patients were informed about the study, and also hospital workers and undergraduate and graduate students. A questionnaire with inclusion and exclusion criteria was administered to those who decided to enter the study voluntarily (see the supplementary material), and those who met the eligibility criteria were provided with the informed consent form for completion.
The variables investigated were latency, amplitude and velocity of the median, ulnar, tibial and peroneal motor nerves, and of the median, ulnar and sural sensory nerves. Latencies of the F-wave and H-reflex were also measured. The variables were analyzed with the Stata 14 software. After obtaining the measures of central tendency of the variables, a Shapiro-Wilk test was carried out to establish if the variables have a normal distribution, for subsequent analysis.
Nerve conduction studies were performed with Nicolet VikingQuest electrodiagnostic system, Nicolet S403 electrical stimulators, gold cup and ring surface recording electrodes, at an average room temperature of 18-20 °C.
Filters were set at 1 Hz-10 KHz, gain of 5 mV per division, electrical pulse duration in 200 ms, and current between 20 and 50 mA for motor nerve conductions, and 20 Hz-3 KHz, gain of 20 μV, electrical pulse duration between 100 and 200 ms, and current between 15 and 30 mA for sensory nerve conductions. The earth pole earth electrode is located on the dorsum of the hand and the dorsum of the foot, respectively.
The active recording electrode (also called G1) is placed in the center of the muscle belly, over the motor plate, and the reference electrode (also called G2) is placed distal to the active electrode, on the tendon of the muscle. The stimulator is placed over the nerve that innervates the muscle, with the cathode located on the side of the active electrode. For the motor nerve conductions, a stimulus was performed at a distance of 8 cm from the G1 electrode, and for the sensory nerve conductions at a distance of 14 cm.
For the motor nerve conduction of the median nerve, the G1 electrode was placed on the belly of the
For the sensory nerve conduction of the median nerve, the G1 ring electrode was placed on the second finger, and G2 electrode was placed 4 cm distal to the G1 electrode. The stimulus was performed 14 cm proximal to the G1 electrode, and posterior to the
For the motor nerve conduction of the ulnar nerve, the G1 electrode was placed on the belly of the
For sensory nerve conduction of the ulnar nerve, the G1 ring electrode was placed on the fifth finger; the G2 electrode was placed 4 cm distal to the G1 electrode. The ulnar nerve stimulus was performed 14 cm proximal to the G1 electrode, and posterior to the
For the motor nerve conduction of the tibial nerve, the G1 electrode was placed over the
For the motor nerve conduction of the peroneal nerve, the G1 electrode is located over the midpoint of the extensor digitorum brevis muscle, on the dorsum of the foot, and the G2 electrode is located distal to the fifth metatarsophalangeal joint. The distal stimulation site is 8 cm proximal to the G1 electrode, lateral to the anterior tibial tendon, and the proximal one was located posterior and inferior to the fibular head.
For the sensory nerve conduction of the sural nerve, the G1 electrode was located behind the lateral malleolus, and the G2 electrode was located distal to it. The stimulation site was located 14 cm proximal to the G1 electrode, lateral to the midline at the back and lower portion of the leg.
The F-wave is a late motor response that occurs after the compound muscle action potential. In the lower limbs, when the peroneal or tibial nerves are stimulated, the G1 and G2 electrodes are placed in the same way as in the motor nerve conductions. The F-wave normally occurs at a latency of 45 to 56 ms. The gain was set to 200 or 500 μV, and the sweep time to 5 or 10 ms, depending on the length of the nerve. Supramaxial stimulation was performed, and the direction of the stimulator was reversed so that the cathode was proximal. The stimulus speed was 0.5 Hz, in order to avoid the effects of the previous stimulus on a later response. Ten records were obtained for each F-wave.
The H-reflex is a true reflex, with sensory afference, synapses and motor efference. It is obtained only by stimulating the tibial nerve in the popliteal fossa and registering in the soleus muscle. The gain was set from 200 to 500 μV. Typical latency is approximately 30 ms. The sweep time was 10 ms, with a stimulus duration of 1 ms.
The G1 electrode was placed over the proximal part of the soleus muscle (where it meets the two gastrocnemius muscle bellies); the G2 reference electrode was placed over the Achilles tendon. The stimulus was performed in the popliteal fossa with the cathode located towards the proximal area, and starting with a low stimulus intensity. It should be stimulated at a rate not faster than once every 2 s (0.5 Hz), in order to avoid the effects of the previous stimulus on a later response. The H-reflex is thriphasic; it appears with latencies of 25 to 34 ms, and is recorded with the muscle at rest.
77 patients, 31 men and 46 women, aged between 18 and 65 years were included in the study. This represents an analysis of 154 nerves for motor and sensory nerve conduction of the 4 extremities, and 142 analyses of late responses (F-wave and H-reflex). The mean age was 32.3 years, and the mean total height was 167 cm (173 cm in men and 163 cm in women).
The mean, standard deviation, and 15th and 75th percentiles for each parameter of motor and sensory nerve conductions and late responses, are shown in tables
Mean ± 2 SD. 15th and 75th percentiles for the total sample.
Mean ± 2 SD. 15th and 75th percentiles for the total sample. The mean is shown, without SD, since variables do not follow normal distribution.
Mean ± 2 SD. 15th and 75th percentiles for the total sample.
Variables were analyzed by groups between men and women and a statistically significant difference was found in the amplitude parameters of the tibial nerve (H: 13 ± 7.2; M: 15 ± 10; p = 0.0099), being greater in women, and in peroneal nerve latency (H: 3.6 ± 1.4; M: 3.3 ± 1.4; p = 0.0091) being greater in men.
To define the reference value of the variables that followed a normal distribution, we took the values below the upper limit of the range established with the standard deviation (2 SD) for latency, and the values above the lower limit of the range established with the standard deviation (2 SD) for amplitude and velocity (
This value corresponds to the 15th percentile since variables do not follow normal distribution
In this study, motor and sensory nerve conductions and late responses were performed in asymptomatic people between 18 and 65 years old, to characterize a young representative sample of the population of Hospital Universitario San Ignacio, in order to establish normality values for the Electrodiagnostic Laboratory.
Latency, amplitude and velocity were analyzed for motor nerve conduction of the median, ulnar, tibial and peroneal nerves, and for sensory nerve conduction of the median, ulnar and sural nerves, and latency of the late responses (F-wave and H-reflex). Normality values were determined according to the parameter to be analyzed. For latency, values below the upper limit of the range are defined as normal, since finding it prolonged would imply myelin damage. For amplitude and velocity, values above the lower limit of the range are defined as normal, because finding it diminished would imply axonal or myelin damage, respectively. Normality values were obtained as established for our laboratory population, shown in
The reference values obtained are similar to those of international studies found in the literature (
The group analysis determined a statistically significant difference in tibial nerve amplitude and peroneal nerve latency, which could be directly influenced by the difference in height among the groups, with the total mean being 167 cm, greater in men than in women (173 cm and 163 cm, respectively).
The results obtained for the reference values can be applied to the patient population studied at the Electrodiagnostic Laboratory of Hospital Universitario San Ignacio. One of the limitations in comparing our results with those of different international studies is the difference in the recording techniques, given the great diversity of methods used, the difference in ages of the subjects analyzed, the different sample sizes, the techniques of data analysis and the generation of normality ranges.
We established reference parameters for motor and sensory nerve conductions of the median, ulnar, tibial and peroneal motor nerves and the median, ulnar and sural sensory nerves, and late responses (F-wave and H-reflex) for the Electrodiagnostic Laboratory of Hospital Universitario San Ignacio. Most of these parameters correlate with the current reference data, with some differences that could be related to the height, sex and characteristics of our population. However, further studies are required to establish these differences.
The authors declare no conflict of interest.
To Camilo García-Pardo, from the Faculty of Economics of the Pontificia Universidad Javeriana.
Female_______ Male_______ ID card number: _______________________ Age: ________ Height (in cm): _________ Contact: ______________________________________
Age between 18 and 65 years: ___________ You are Colombian: ___________
Tingling symptoms: _____________ Lack of feeling: _____________ Burning or electric pain in extremities: _____________ Known peripheral neuropathy: _____________ Hereditary (familial) neuropathy: _____________ Compressive neuropathy (e.g., carpal tunnel syndrome): _______________ Autoimmune diseases: _____________ Spinal diseases or radiculopathies: _____________ Metabolic diseases: _____________ History of cancer/chemo use: _______________ Neuromuscular diseases (e.g., myasthenia gravis): _______________ Amputation of a limb: _______________ Contraindication for supine position: _______________